beautypg.com

OWON MSO Series User Manual

Page 92

background image

87

Fig.5-7

5:Search trigger position

2. Search specified value in BUS
Searching BUS menu as below:

Function

Setting

Instruction

BUS

BUS0~BUS3

Choose searching BUS

Code type

0x0000~0xffff

(HEX)

0~65535(DEC)

Can be set discretionarily between 0x0000 and
0xffff (HEX)or between 0 and 65535 (DEC)
according to the bus and code setting.

Previous

Search the matched value prior to current cursor

Next

Search the matched value after current cursor.

For example, follow up below steps to search the data of 0x18 values in BUS0

A、Press "2(SEARCH)" and menu appears.
B、Press "F1" till target display as "BUS".
C、Press "F2" till BUS SEL as "BUS0".
D、Press "F3" and choose "Code Type ", value under displayed with red background

and green digital indicator in the panel is on. Then insert the value of "18" by
digital key in the panel to get start.

E、Press "F3" again and red background for value under code is disappeared and

value setting finished. Repeat operation of steps D. E if need to modify the value.

F、Press "F4" and choose "Previous" to searching the matched value prior to current

cursor. Cursor will stop in this value if it has and info window show "Got the
target
" and if it hasn't it will show "Search failed".

G、Press "F5" and choose " Next" to searching the matched value after current cursor.