beautypg.com

System status (u0) return codes – Measurement Computing Digital488/80A User Manual

Page 76

background image

System Status (U0) Return Codes

C#

Configuration

C0

All 5 ports are input.

C1

Port 1 is output; the other ports are input.

C2

Ports 1 and 2 are output; the other ports are input.

C3

Ports 1 through 3 are output; the other ports are input.

C4

Ports 1 through 4 are output; Port 5 is input.

C5

All 5 ports are output.

E#

Query Error Status

E0

No Error.

E1

Unrecognized Command.

E2

Invalid Command Parameter.

E3

Command Conflict Error.

E5

Non-Volatile RAM Checksum Failure.

E6

Internal Data Buffer Overrun

Note: Except for

E5

, error is cleared upon reading error status.

F#

Data Format

F0

ASCII hexadecimal (4 bits per character).

F1

ASCII character (4 bits per character).

F2

ASCII binary (1 bit per character).

F3

ASCII decimal (8 bits per number).

F4

Binary (each byte = 8 bits).

Note: In Data Format

F5

(high-speed binary), the command interpreter is disabled.

G#

Bus Input/Output

G0

Input and output port data are sent when addressed to Talk.

G1

Only input port data is sent

G2

Only output port data is sent

G3

Buffered input and output data are sent

G4

Continuous buffered input and output data are sent

I###

Invert

I0

All handshaking/control lines are active high, all data lines are high true.

I1

The Inhibit output line is active low.

I2

The Trigger output line is active low.

I4

The Data Strobe output line is active low.

I8

The Clear output line is active low.

I16

The Data lines are low true.

I32

The EDR input line is falling-edge sensitive.

I64

The Service input line is falling-edge sensitive.

Note:

The status message reflects the sum of all received Invert commands.

70 Digital488/80A Commands

967695

Digital488/80A User's Manual