CMG-DCM Data Communications Module

This option has the same effect as the MS-GAP command on a DM24.

If the port is not running the

gcf_out service, this setting will be

ignored.

Configure – Digitizer

Clicking on

Configure – Digitizer opens a page allowing you to

reconfigure the digitizer attached to the port, if there is one.

When the page loads, the DCM attempts to retrieve the current

configuration of the digitizer attached to that port. This will take a few
seconds, after which the message Configuration successfully
retrieved from attached instrument should appear together

with a form detailing the available settings.

If the DCM cannot find a digitizer attached to the port, it will give the

message Unable to retrieve configuration from attached
instrument. In this case you should

check that an instrument is connected to the port you clicked,

and that it is powered up;

check that the baud rate of the serial port (see Section 6.2, page
82) matches the output baud rate of the digitizer; and

check that the DCM is running the

gcf_in service on the port.

For a full description of the digitizer settings you can change, see

Chapter 7, page 105.

6.3 Disk

On this page you can set how the DCM uses its hard disk, and what

data it records.

86

Issue F