beautypg.com

Guralp Systems CMG-DCM User Manual

Page 42

background image

CMG-DCM Data Communications Module

If the digitizer is not responding at all, check the power supply

to the digitizer and the cable between it and the DCM.

The DCM is receiving streams, but they do not contain any data.

If the data you see is zero:

Increase the Y axis scale using the

icon in

Scream!'s

Waveview window or the corresponding icon in the

DCM's

Data Viewer.

Check the mass position outputs to see if the masses are locked.

A properly unlocked and centred mass should show a mass
position within 1000 counts of zero; a locked mass will give a

value greater than 32,000 (or less than −32,000) counts. If your
sensor has remote mass locking, you can unlock the masses by

navigating to Actions Digitizer setup

and clicking

Unlock

sensor.

Locking or unlocking the sensor mass typically takes several

minutes to complete.

Check the connection between the sensor and digitizer and try
again.

If you cannot see the data in the stream, remove any constant offsets
by clicking on the Zero streams icon

in Scream!'s

Waveview

window or the DCM's Data Viewer.

The DCM receives streams, but gaps appear in the data some
minutes after boot-up.

Check that the baud rate between the digitizer and the DCM is

sufficient for all the data streams you want to transmit. If it is not, the
digitizer's output buffer will gradually fill up until no more data can be

stored. Increase the baud rate of the digitizer through the DCM (or
using Scream!),

then set the baud rate of the DCM's input port to the

same value.

If you are using triggered output streams, be especially careful to allow
a high enough baud rate to transmit data from all possible output

streams simultaneously, or you will observe gaps when an event
triggers the digitizer.

42

Issue F