beautypg.com

Sensitivity:correction:offset[:magnitude, Sensitivity:frequency – EXFO PM-1100 Optical Power Meter User Manual

Page 99

background image

Remote Control Commands

Optical Power Meter

89

Product-Specific Commands—Description

SENSitivity:CORRection:OFFSet[:MAGNitude]

Description This command sets an offset value that is applied when dB or

dBm units are selected.

Syntax

SENS:CORR:OFFS[:MAGN]
[DB]

Parameters The is an offset with dB units in the ±99.999

format.

Example

SENS:CORR:OFFS:MAGN 22.105

SENSitivity:FREQuency

Description This command selects a data sampling rate.

Syntax

SENS:FREQ

Parameters The is the sampling rate with sec

-1

units.

0.1, 0.5, 1, 5, 10, 20, 40 samples per second.

Example

SENS:FREQ 20

SENSitivity:FREQuency?

Description This query returns the current sampling rate.

Syntax

SENS:FREQ?

Response

Returns the current sampling rate in samples/second units.

Example

SENS:FREQ?