beautypg.com

EXFO PM-1100 Optical Power Meter User Manual

Page 81

background image

Remote Control Commands

Optical Power Meter

71

IEEE 488.2 Commands—Quick Reference

*RST

Description This command empties the step response list. It is only seen

when it is part of another multiple command. In the example

below, by adding this command after *IDN?, you will not be

able to access the answer. The *RST, in this instance, erases the

identification string. In addition, this command performs the

following operations:

1. Return to initial state before command was sent, and not

necessarily to previous settings.
2. Force the device to enter into an Operation Complete

Command Idle State (OCIS).
3. Force the device to enter into an Operation Complete

Query Active State (OQAS).
4. Initialize previous responses unless there has been a

program message terminator preceded by an *RST.

Syntax

*RST

Example

*IDN?;*RST