Remote Control Commands

80

PM-1100

Product-Specific Commands—Description

INITiate:CONTinuous

Description This command starts or stops continuous mode. When

continuous mode is activated, the latest value is continuously

stored in the initialized memory location so it can be fetched at

any time with the FETC[:SCAL]:POW:DC? query. Values can be

power or loss measurements, depending on the current

measurement mode of the PM-1100.

Syntax

INIT:CONT

Parameters The value can be

“0” to stop continuous mode
“1” to start continuous mode

Example

INIT:CONT 1

Note

Use the INIT:STOR command to initialize a memory location.

See also

FETC:SCAL:POW:DC?, INIT:CONT?, INIT:IMM, INIT:STOR, and

READ:SCAL:POW:DC?

INITiate:CONTinuous?

Description This query returns a value indicating whether continuous mode

is activated.

Syntax

INIT:CONT?

Response

“0” if continuous mode is stopped
“1” if continuous mode is started

Example

INIT:CONT?

See also

FETC:SCAL:POW:DC?, INIT:CONT, INIT:IMM, and