beautypg.com

EXFO PM-1100 Optical Power Meter User Manual

Page 79

background image

Remote Control Commands

Optical Power Meter

69

IEEE 488.2 Commands—Quick Reference

*ESR?

Description This query reads the contents of the Standard Event Status

Register (ESR).

Syntax

*ESR?

Response

A binary integer between 0 and 255.

*IDN?

Description This query reads the IQ system identification string.

Syntax

*IDN?

Response

“EXFO E.-O. Eng PM-1100 Vxx.xx”, where xx.xx is the current

product version.

*LOK

Description This command is used to set the Remote Lockout programming

state.

Syntax

*LOK

Notes

This command can only be used when working with RS-232

communication.