beautypg.com

Product-specific commands-description, Product-specific commands—description, Abort – EXFO PM-1100 Optical Power Meter User Manual

Page 87: Display:dimmer

background image

Remote Control Commands

Optical Power Meter

77

Product-Specific Commands—Description

Product-Specific Commands—Description

ABORt

Description This command stops any measurement in progress.

Syntax

ABOR(0..26)

Example

ABOR3

DISPlay:DIMMer

Description This command is used to adjust the intensity of the PM-1100

display (high or low) or to turn off the display without turning off

the unit.

Syntax

DISP:DIMM

Parameters The parameter can be “HI”, “LO”, or “OFF”.

Example

DISP:DIMM OFF

Note

When the display is turned off by this command, any key

pressed on the PM-1100 keypad will return the display to high

intensity. To prevent this, lock the PM-1100 keypad with the

*LOK command.

See also

DISP:DIMM?, *LOK, and *LOK?