beautypg.com

Elength – EXFO PM-1100 Optical Power Meter User Manual

Page 94

background image

Remote Control Commands

84

PM-1100

Product-Specific Commands—Description

MMEMory:ACQuisition:DATA:RECAll:VALUE?

Description This command returns the measurement saved in the

memory location.

Syntax

MMEM:ACQ:DATA:RECA:VALUE?

Parameters The parameter represents the memory location from

which to recall the measurement data (out of the 1024

available).

Response

Measurement data saved in the specified memory location in

the “±999.999E±99” format.

Note

Use the MMEM:ACQ:DATA:RECA:UNIT? query to get the

corresponding measurement units.
Use the MMEM:ACQ:DATA:RECA:WAV? query to get the

wavelength at which the measurement was taken.

Example

MMEM:ACQ:DATA:RECA:VALUE? 1021

MMEMory:ACQuisition:DATA:RECAll:WAV

elength

?

Description This query returns the wavelength corresponding to the last

value fetched with the MMEM:ACQ:DATA:RECA:VALUE? query.

Syntax

MMEM:ACQ:DATA:RECA:WAV?

Response

A wavelength in the “9999 nm” format.

Example

MMEM:ACQ:DATA:RECA:WAV?