95

C SCPI-Based Errors

Error

Number

Description

Probable Cause

-100

“Command error.”

An error occurred while validating a

command.

-101

“Undefined header.”

Unknown command.

-102

“Missing parameter.”

A command parameter is missing.

-103

“Parameter not allowed.”

An extra parameter is present.

-104

“Data type error.”

Invalid parameter format.

-200

“Execution error.”

An error occurred while executing a

command.

-300

“Device dependent error.”

A command has taken longer than

expected to complete execution.

-300

“Light on detector.”

Light reached the detector during

the offset nulling.

-300

“Invalid wavelength.”

The entered wavelength is out of

range.

-300

“Parameter out of range.”

The command parameter is out of

range.

-300

“No stored data.”

No data is stored in the specified

location.

-300

“Program running.”

An acquisition is already in progress.

-300

“Program duration not

initialized.”

No duration parameter is currently

defined for the program.

-300

“No samples available.”

No sample parameter is currently

defined for the program.

-300

“No data.”

No data to recall.

-300

“Invalid unit.”

The requested measurement units

are invalid.

-300

“Invalid resolution.”

The requested resolution is invalid.