EXFO FTB-610 Wideband Copper Test Module User Manual

Page 221

background image

RFL Test

Copper Test

213

Setting Cable Parameters

The page allows you to set the following parameters:

No. of Cable Sections allows you to set the number of consecutive
cable sections, to a maximum of 5. If unknown, set the number to 1.

Section No. lists the number of cable sections.

Length allows you to specify the cable length of each section. If the
cable length is unknown, specify 0 and the RFL test will measure the
length.

Cable Type allows you to select a cable entry (with its associated
parameters) from the current Cable Book, to use for testing. The
selection is not saved when leaving the test. The current Cable Book
entry is shown by default.

Temperature allows you to change the temperature of the cable under
test in either degree C or F.

Resistance specifies a value for the resistance constant of the cable in
each section.

Load Coil allows you to set the detection of the presence of load coils
by checking the box.