Pair Detective Test

92

FTB-610

Pair Detective

Setting Threshold Values

The Test Thresholds page allows you to set the Pair Detective threshold
values.

To set the Threshold values:
1.
From the Copper main menu Test Groups, select Pair Detective test

.

2. From the Tests menu, select Pair Detective test.

3. Tap Edit Thresholds button to open the Test Thresholds page.

4. Configure the thresholds as required.

5. Tap Close to confirm and exit the page.

The page allows you to set the following parameters:

Foreign DC Voltage sets the maximum foreign DC voltage value.

POTS DC Voltage sets the minimum working POTS DC voltage value.

POTS DC Current sets the minimum working POTS DC current value.

Capacitive Balance sets the minimum capacitive balance threshold as
a percentage.

Good Balance sets the good balance threshold.