71M6541 Demo Board REV 3.0 User’s Manual

25

Rev 4.0

v_limit

Error if exceeded.*

Same units as CE’s v0sqsum.

407.3V =
240V*sqrt(2)
*120%

)6

signed

32

wrate_mpu

Convert from CE
counts to pulses.

CE’s w0sum units per pulse,
rounded up to next largest CE
count so Wh accumulation and
display is always rounded down.

3.2 Wh for 3-
phase
1.0 Wh for 1-
phase

)7

signed

32

interval

The number of
minutes of a de-
mand interval.

Count of minutes.
(60/interval)*interval = 60.

2 minutes.

)8

unsigned

8

mains_hz

Expected number of
cycles per second
of mains. 0 disables
the software RTC
run from mains.

Hz

0

)9

unsigned

8

temp_cal1

Machine-readable
units per 0.1C

Refer to the IC data sheet.
Temperature is calculated as
temp = (measured_temp –
temp_datum)/temp_cal1 +
temp_cal0

Refer to the IC
data sheet.

)A

signed

32

mtr_cal1
[0..3]

9

Linear temperature
calibration for meter
elements A..D.

ppm*(T -

mtr_datum), in 0.1˚C

150

)B..E

signed

16

mtr_cal2
[0..3]

9

Squared tempera-
ture calibration for
meter elements
A..D.

ppm2*(T -

mtr_datum)2, in 0.1˚C -392

)F..1

2

signed

16

y_datum

Center temperature
of the crystal.

0.1C

25C

)13

signed

16

y_cal1

5

RTC adjust, linear
by temp.

10ppb*(T -

y_datum), in 0.1˚C

0

)14

signed

16

y_cal2

5

RTC adjust,
squared by temp.

1ppb*(T -

y_datum)2, in 0.1˚C

38

)15

signed

16

s_cal

1

Accumulation inter-
vals of
Autocalibration

Count of accumulation intervals
of calibration.

accumulation
intervals cover
both chop polar-
ities.

)16

signed

16

v_cal

1

Volts of
Autocalibration

0.1V rms of AC signal applied to
all elements during calibration.

2400
240 V is a
standard full-
scale setup for
meter test.

)17

signed

16

i_cal

1

Amps of
Autocalibration

0.1A rms of AC signal applied to
all elements during calibration.
Power factor of calibration signal
must be 1.

300
30 A is a stand-
ard full-scale
setup for meter
test.

)18

signed

16