u To zero adjust a custom probe

This procedure zero adjusts a custom probe and sets its intercept value based on an actual
sample using the applicable custom probe.

(1) Connect the EA-200 to your ClassPad, and connect the custom probe you want to zero

adjust to the EA-200’s CH1 channel.

(2) What you should do next depends on whether you are calibrating a new custom probe

or an existing custom probe whose settings have been edited.

When calibrating a new custom probe:
• Perform steps (1) through (5) under “To configure a new custom probe” on page 3-12.
• You can skip step (2), if you want.

When calibrating an existing custom probe:
• Perform steps (1) through (6) under “To edit an existing custom probe” on page 3-14.

(3) Tap [Zero Adjust].

• This starts sampling by the sensor connected to the

CH1 channel and displays a Sampling... dialog box
like the one shown to the right. This is how the dialog
box appears during standby prior to the sample.

• If this dialog box is left on the display, sampling will

terminate and the dialog box will close automatically
after about five hours.

(4) Tap [Set] when you are ready to take the sample.

• This closes the Sampling... dialog box and returns to the Custom Probe dialog box.

• The value produced by the sample is entered automatically for [Intercept].

(5) After you are finished, tap [Set] to save the custom probe and close the Custom Probe

dialog box.

Tip

• An error occurs if the sample value in step (4) is not correct. If this happens, repeat the procedure

from step (3).