3-17

Setup

(5) In the [Value 1] box, input the reference value for the

first sample, and then tap [OK].

• This restarts sampling by the sensor connected to the

CH1 channel and displays a Sampling... dialog box
like the one shown to the right. This is how the dialog
box appears during standby prior to the second
sample.

• If this dialog box is left on the display, sampling will terminate and the dialog box will

close automatically after about five hours.

(6) Tap [Set] when you are ready to take the second sample.

• This causes the Sampling... dialog box to change as

shown to the right.

(7) In the [Value 2] box, input the reference value for the second sample, and then tap

[OK].

• This closes the Input Value dialog box and returns to the Custom Probe dialog box.

• The EA-200 calculates the slope and the intercept for the linear interpolation formula

from the first set of values (reference and sampled), and the second set of values,
and automatically inputs them into the [Slope] and [Intercept] values of the custom
probe.

• After you are finished, tap [Set] to save the custom probe and close the Custom

Probe dialog box.

Tip

• An error occurs if the reference values you input in steps (4) and (6) are not correct. If this

happens, repeat the procedure from step (3).