Stream mode, Preparation and performance (examples) – A&D Weighing HW-200KV-WP User Manual
Page 53
HV-WP/HW-WP Series
Page 51
16. RS-232C Serial Interface
16.2. Stream Mode
The scale outputs the current display data at the time when refreshing the display.
The scale does not output data while in the setting mode.
Averaging of function table
Refresh rate
A
f13 0 Fast
response
Sensitive value
Approximately 7 times/s while the display is unstable,
Approximately 4 times/s when the display is stable
A
f13 1
Ú
A
f13 2 Slow
response
Stable
value
Approximately 4 times/s
A
16.2.1. Preparation and Performance (Examples)
Step 1 Turn off the display.
Press
the
ON/OFF
switch while the
ZERO
switch is
pressed and held. The function table is displayed.
Step 2 Press the
ENTER
switch to display f4 ("Baud
rate"). Select a parameter of "Baud rate" with the
∧
switch.
Step 3 Press the
ENTER
switch to display f5 ("Output
mode").
Select "Stream mode ( f5 0 )" with the
∧
switch.
Step 4 Press the
ENTER
switch to display f15 ("Format").
Select "Format 1 ( f15 0 )" with the
∧
switch.
Step 5 Press the
ENTER
switch to display end .
Step 6 Press the
ENTER
switch to store the new
parameters. The scale returns to the normal
weighing mode and outputs the data continuously.
P
ress and hold
Press
S T , 0 0 0 0 . 0 0 0 k g
U S , 0 0 1 0 . 6 4 5 k g
S T , 0 0 1 0 . 6 4 4 k g