beautypg.com

Accumulation function – A&D Weighing HW-200KV-WP User Manual

Page 28

background image

Page 26

HV-WP/HW-WP Series

10. Accumulation Function

10. Accumulation Function

This function counts the number of weighed items, calculates the total mass value

and can display the number and accumulated mass value.

The accumulation function is displayed with up to 6 digits. The balance can not display 7 or

more digits, therefore the leading digits are not displayed.
Example: With 60K type, when importing the data of 17 accumulations of up to 60kg (60.000

X 17=1020.000), the balance displays this as “020.000”.

Set the parameters of the "accumulation function (

f8

)" in the function table, in

advance, to use this function.

The accumulation count and accumulated mass value are stored in the scale even if

the power is removed.

Operation and Switches

The display of the accumulation count has a blinking without a unit.

The display of the accumulated mass value has a unit and a blinking .

Pressing the

MODE

switch, the accumulation count and accumulated mass value are

displayed.

Pressing

the

ZERO

switch in the accumulation function ( is blinking), the current

function resets. (The count and accumulated mass value become zero.)

Caution

The accumulation function can be used with the first weighing unit accumulated.

Parameter List and Word Definition

The "nearly-zero" is within ±5 digits from the zero point in the unit of kg.
The "digit", a unit of display, is equivalent to the minimum measurable mass.
The "zero point" is the fundamental starting point to weigh anything.

Function table

Meaning and purpose

A

f8 0

Accumulation function not used.

A

f8 1

The scale accumulates the data, if the

F

switch is pressed, when

the display is a positive stable value without nearly-zero. The next
accumulation can be performed after the display is nearly-zero or a
negative value.

A
A
A
A

f8 2

The scale accumulates the data, if the

F

switch is pressed, when

the display is a stable value and without nearly-zero. The next
accumulation can be performed after the display is nearly-zero.

A
A
A


f8 3

When the display is a positive stable value and without nearly-zero,
the scale accumulates the data automatically. The next
accumulation can be performed after the display is nearly-zero or a
negative value.