14. Calibration (Adjusting Scale)

14.2. The Complete Calibration Procedure

14.2.1. Gravity Acceleration Correction

Step 1 Turn on the display.

Open the rear cover of the display unit.

Press and hold the

CAL

switch, it is in the depth of

5cm, to enter the calibration mode.

Then

Cal 0

is displayed.





Step 2 Press the

ZERO

switch to enter the gravity

acceleration correction mode.



Step 3 Set your local gravity acceleration using the following

switches.

switch

Selecting the number of a figure.

switch

Selecting a figure.



Step 4 Press the

ENTER

switch to store the new value.

Then

Cal 0

is displayed.



Step 5 Press the

CAL

switch again. Then the scale returns to

the normal weighing mode.


14.2.2. Preparation

Step 6 Confirm the environmental conditions as follows:

Maintain a constant temperature and stable power.

Install the scale on a solid floor where there is no draft, vibration, strong magetic

fields or direct sunlight.

Consider section "4. Caution".


Step 7 Display normal weighing for at least 30 minutes to warm up the scale.