11. Upper/Lower Comparator Function

Function table

Meaning and purpose

A

f6 7

When the display value becomes stable and not nearly-zero, the
scale compares the value. It does not compare on an unstable
condition.

A
A

Caution

The upper limit value (HI) must be greater than the lower limit value (LO).

The parameters of the upper limit value (HI) and the final value (HI) use the
same memory. The parameters of the lower limit value (LO) and the
preliminary value (OK) use the same memory.

The upper/lower comparator function, the simple batch function and the

full/dribble batch function can not be used at the same time because these
parameters use common memory.



11.1. Preparation (Setting Parameters)

Step 1 Turn off the display.

Press

the

ON/OFF

switch while the

ZERO

switch is

pressed and held. The function table is displayed.





Step 2 Press the

ENTER

switch to display an item of the

comparator function (

f6

).



Step 3 Select a parameter of the comparator function

(

f6 0

~

f6 7

) with the

switch.



Step 4 Press the

ENTER

switch and the

F

switch to display

end . Press the the

ENTER

switch to return to the

normal weighing mode.