A&D Weighing HJ-150 User Manual
Hj-150 compact scale operating instructions, Cautions, Specifications
1
HJ-150 Compact Scale
Operating Instructions
Cautions
Avoid installing / leaving the scale in direct sunlight / high humidity.
Place the scale on a flat and solid surface in a place free from wind.
Avoid vibration, strong electric or magnetic fields and rapid changes in
temperature or humidity.
Avoid using the scale near other electronic equipment.
Do not mix the battery types. Replace all of the batteries at same time.
If the scale is not to be used for a long period of time, remove the batteries.
Avoid over / shock load that could cause damage to the scale.
Keep the protective cover attached when you carry the scale.
Wait for more than 6 seconds to let the scale warm up after turning on
the power.
When the location to be used or the ambient environment has changed,
calibrate the scale before use.
When the display shows lo , or continues to show 8.8.8.8 more than
30 seconds after turnig on, replace the batteries with new one.
Specifications
Capacity: 150 x 0.1 g / 5.29 x 0.01 oz
Linearity: ±0.2 g Repeatability: 0.1 g (Standard deviation)
Operating environment: 5
○
C ~ 40
○
C (41
○
F ~ 104
○
F), less than 85% RH
Display: 7 segment LCD with backlight (Character height 8 mm)
Power: 2 x “AAA” size batteries, approx. 80 hours with alkaline batteries.
Pan size: 60 (W)x 64 (D) mm / 2.36 (W) x 2.52 (D) in.
Accessories : Instruction manual, 100g weight, 2 monitor batteries
LCD
LIGHT/CAL
Weighing pan
ON/ZERO/OFF
Battery cover
g / oz
Open
Battery
+