14. Calibration (Adjusting Scale)

14.2.3. Calibration of the Zero Point

Step 8 Press and hold the

CAL

switch to enter

the calibration mode after displaying
normal weighing for 30 minutes.

Then

Cal 0

is displayed.



Step 9 With nothing on the weighing pan, press

the

ENTER

switch while the stable mark is

displayed. The scale stores the current
condition as the zero point.



Step 10 The scale displays

5pn1

for several

seconds. Finishing the calibration mode
at this stage, proceed to step 14.


14.2.4. Span

Calibration

Step 11 Set the value of the calibration mass

using the following switches. (This initial
value is according to each model.)

switch

Selecting the number of a
figure.

switch

Selecting a figure.


Step 12 Place the mass on the weighing pan

which was set at step 11, press the

ENTER

switch while the stable mark is

displayed. The scale then calculates the
span and stores it.


Step 13 The scale displays

end

at the finish.

Remove the mass from the weighing pan.



Step 14 Press the

CAL

switch to return to the normal

weighing mode.

Note
The pan shape depends
on the scale model.

For several seconds

Mass value

15kg

CAL switch

STABLE

STABLE

Stability mark

,

STABLE

Stability mark

Weighing value

CAL switch

Weighing value