Specifications – A&D Weighing GF-32K User Manual
Page 93
91
20. Specifications
GX-8K GX-8K2 GX-10K GX-12K GX-20K GX-30K GX-32K
Weighing capacity
8.1 kg
10.1 kg
12 kg
21 kg
31 kg
31 kg
8.1008 kg
31.008 kg
Maximum display
8.10084 kg
2.10009 kg *1
10.10084 kg 12.0084 kg 21.0084 kg 31.0084 kg
6.1009 kg *1
Minimum weighing value (1 digit) 0.01 g 0.1 g / 0.01 g
0.01 g
0.1 g
1g / 0.1 g
Repeatability
(Standard deviation)
0.01 g
0.05 g /
0.01g
0.01 g
0.1 g
0.5g / 0.1 g
Linearity
±0.03 g
±0.1 g /
±0.02 g
±0.03 g
±0.2 g
±1g / ±0.2 g
Stabilization time
(Typical
at
FAST
)
Approx. 1.5 seconds
Sensitivity drift,
*2
(10°C ~ 30°C / 50
°F ~ 86°F) ±2 ppm/°C ±3 ppm/°C ±2 ppm/°C
±3 ppm/°C
±5 ppm/°C
Accuracy right after
calibration using the
internal mass
(Accuracy of full scale) *3
±0.15 g
±0.3 g
±0.15 g
±1.0 g
±1.5 g
±3 g
Operating environment
5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)
Internal mass
Built-in function
Weighing data of data
memory
200 data, 100 data with calendar
Time and clock function
Built-in function
Display refresh rate
5 times/second or 10 times/second
Minimum unit
mass
0.01 g
0.1 g
0.01 g
0.1 g
1g
Counting
mode Number of
samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 g
10 g
1 g
10 g
100g
Percent mode
Minimum 100%
display
0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)
Interface
(Provided as standard)
RS-232C with Windows Communication Tools Software WinCT
External calibration weight
2 kg, 3 kg, 4 kg, 5 kg,
6 kg, 7 kg, 8 kg
2 kg, 3 kg,
4 kg, 5 kg,
6 kg, 7 kg,
8 kg, 9 kg,
10 kg
5 kg, 10 kg 10 kg, 20 kg
20 kg, 30 kg
Weighing pan
270 x 210 mm
External dimensions
300(W) x 355(D) x 111(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 11VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power receptacle type.
Weight
Approx. 9.3kg
Dust and water protection
Complying with IP65
*1: The balance allows weighing using a precision range, even with a heavy tare placed on the pan. (Smart range function)
*2: When automatic self calibration is not used.
*3: Accuracy right after calibration using the internal mass in good ambient conditions (within the temperature range of
10
°C to 30°C (50°F to 86°F) with no abrupt changes in temperature or humidity, no drafts, no effect by magnetic fields
or static electricity).
Check the internal mass periodically as described in "18. Maintenance".