12-5-2. Recalling the upper and lower limit values

(Quick selection mode)

The procedure below describes an easy way to recall the upper and lower limit values to be
used for weighing. When the recalled upper and lower limit values are to be changed, refer
to "12-5-1. Storing the upper and lower limit values".


Note

Refer to "12-5-1. Storing the upper and lower limit values" for the method to store the

upper and lower limit values.


Step 1 Set the "Data memory (

data

)" parameter to "

4

".


Step 2 Press the

MODE

key to enter the upper/lower

limit value recalling mode. The upper limit value
last selected with its comparator number appears.
The display is as shown to the right, with all the
digits blinking.


Step 3 Press the

MODE

key to select the value. Each

time the

MODE

key is pressed, the displayed

value changes as follows:

(........

C03

HI

C03

LO

C04

HI

C04

LO

↔ ........)


Step 4 Press the

PRINT

key to confirm the selection. The

balance returns to the weighing mode with the
selected upper and lower limit values ready for use.

To cancel the selection, press the

CAL

key. The

balance returns to the weighing mode.


Note

When no operation is performed in step 4 (after a few seconds of inactivity), the

balance selects the value currently displayed and returns to the weighing mode
automatically.


Comparator

number

Upper

limit

Store

Comparison with

no.

04

Lower

limit

Target

Number

Quick selection