10-8-1. Example

1

Comparison when the weighing data is stable or overloaded, excluding "near zero", upper
limit and lower limit.

Selecting a comparator mode

Step 1 Press and hold the

SAMPLE

key until

ba5fnc

of the function table is displayed,

then release the key.

Step 2 Press the

SAMPLE

key several times to display

Cp fnc

.

Step 3 Press the

PRINT

key.

Step 4 Press the

RE-ZERO

key several times to display

Cp 1

.

Step 5 Press the

SAMPLE

key to display

Cp in

.

Step 6 Press the

RE-ZERO

key several times to display

Cp in 0

.

Step 7 Press the

PRINT

key to store the selected mode.

Entering the upper and lower limit values

Step 8

With

Cp Hi

displayed, press the

PRINT

key. The current setting of the upper

limit value is displayed with all of the digits blinking.

When the current setting is not to be changed, press the

PRINT

or

CAL

key to

go to step 9.

When the current setting is to be changed, press the

RE-ZERO

key.

The balance is now in the digital input mode. To use the weighing input mode,

press and hold the

MODE

key.

Digital input mode

Change the setting using the following keys.

SAMPLE

key.....To select the digit to change the value.

RE-ZERO

key ...To change the value of the digit selected.

MODE

key........To switch the polarity.

PRINT

key........To store the new setting and go to step 9.

CAL

key...........To cancel the new setting and go to step 9.

Weighing input mode

Press

the

RE-ZERO

key. The balance displays

0.0 g

. Place a sample, with a mass

that corresponds to the upper limit value, on the pan. Press the

PRINT

key to

store the upper limit value. Remove the sample. The balance displays

Cp lo

.

Step 9 With

Cp lo

displayed, press the

PRINT

key. The current setting of the lower limit

value is displayed with all of the digits blinking.

When the current setting is not to be changed, press the

PRINT

or

CAL

key to go

to step 10.

When the current setting is to be changed, press the

RE-ZERO

key.

The balance is now in the digital input mode. To use the weighing input mode, press
and hold the

MODE

key.

Enter the lower limit value in the same way as described in step 8. Then, go to step 10.

Step 10 Press the

CAL

key to exit the comparator function and return to the weighing mode.