4. description of the item "data output mode – A&D Weighing GF-32K User Manual
Page 41
39
10-4. Description Of The Item "Data Output Mode"
The parameter setting of "Data output mode (
prt
)" applies to the performance when the
"Data memory (
data
)" parameter is set to "
2
" (to store the weighing data) and when the data
is transmitted using the RS-232C interface.
Key mode
When
the
key is pressed with the stabilization indictor turned on, the balance
outputs or stores the weighing data and the display blinks one time.
Required
setting
dout
prt 0
Key
mode
Auto print modes A and B
When the displayed value is stable and the conditions of "Auto print polarity", "Auto print
difference" and reference value are met, the balance outputs or stores the weighing data.
When
the
key is pressed with the stabilization indictor turned on, the balance
outputs or stores the data and the display blinks one time.
Auto print modes A
Example
For weighing each time a sample is placed and removed, with "
ar-d
"
set to "
1
" (to adjust zero after the data is output).
Required
setting
dout
prt 1
Auto print mode A (reference = zero)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
dout
ar-d 1
Zero after output
Auto print modes B
Example
For weighing while a sample is added.
Required
setting
dout
prt 2
Auto
mode
B (reference = last stable value)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
Stream mode
The balance outputs the weighing data continuously regardless of the display condition. The
display does not blink in this mode. The interval memory mode is used when the "Data
memory (
data
)" parameter is set to "
2
" (to store the weighing data).
Example
For monitoring data on a computer.
Required
setting
dout
prt 3
Stream mode
dout
data 0
Data memory function is not used
ba5fnc 5pd
Display refresh rate
5if
bp5
Baud rate
Caution The balance may not transmit the data completely at the specified refresh rate,
depending on the baud rate or data added to the weighing data such as time,
date and ID number.