10-2. Details Of The Function Table
Class
Item and Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment. With "Hold 1", sets the
averaging time.
0
Stable when within ±1 digit
1
5t-b
Stability band width
2
Stable when within ±3 digits
The stabilization indicator illuminates
with the display fluctuation within the
range. With "Hold 1", sets the stable
range.
0
OFF
Hold
Hold function
1
ON
Holds the display when stable in
animal mode. With "
Hold 1
",
ANIMAL
turns on.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate 1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
0
OFF
poff
Auto display-OFF
1
ON (10 minutes)
Turns off the display after 10
minutes of inactivity.
0
OFF
g5i
Capacity indicator
1
ON
Capacity indicator.
Zero:
0%
Maximum capacity: 100%
0
OFF
add
Accumulation function 1
ON
Displays and outputs the total value
of the weighing data.
0
Does not display
ba5fnc
Environment
Display
rng
Display at start
1
Displays
Select whether or not to display the
smallest displayable weighing value
at weighing start.
Cl adj
Clock
Refer to "10-7. Clock and Calendar Function"
Confirms and sets the time and
date. The time and date are added
to the output data.
0
No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
3
Continuous comparison, excluding "near zero"
Cp
Comparator mode
4
Continuous comparison, including "near zero"
0
Set the upper lower limit value
Select Cp Hi or Cp lo.
Cp in
Data input method
1
Set the reference value
Select Cp ref or Cp lmt.
0
Not added
Cp-r
Comparison results 1
Added
Select whether or not to add the
comparison results to the output
data.
0
OFF
Cp-b
Main display comparison
1
ON
Displays the results on the main
portion of the display in place of the
weight value.
0
OFF
bep
LO buzzer
1
ON
Select whether or not to sound the
LO buzzer.
0
OFF
bep
OK buzzer
1
ON
Select whether or not to sound the
OK buzzer.
0
OFF
Cp fnc
Comparator
bep
HI buzzer
1
ON
Select whether or not to sound the
HI buzzer.
: Functions for GX-K series. : Factory settings. Digit is a unit of minimum weighing value.
Displayed
only when
Comparator
output
(GX-04K) is
installed