8-6. Correcting The Internal Mass Value Of The GX-K Series

The balance can correct the internal mass value within the range shown below. This

function corrects the internal mass value to conform to an external weight. The corrected
mass value is maintained in non-volatile memory even if the AC adapter is removed. The
internal mass value is corrected as follows:

Note GF-K series can not use this function.

Model Target

Range

GX-8K GX-8K2
GX-10K

2kg ±0.50g

GX-12K GX-20K
GX-30K GX-32K

10kg ±5.0g


Step 1 Calibrate the balance using the internal mass. (one-touch calibration).

Place the external weight and find out the correction value.

Example for correcting the weight value by +0.3 g in 10 kg, when

using GX-12K. If correcting the weight value by +0.3 g in 5 kg,
the weight changed into 10 kg, the correction value is +0.6 g.

Step 2 Press

the

ON:OFF

key to turn off the display.

Step 3 While pressing and holding the

PRINT

key and

the

SAMPLE

key, press the

ON:OFF

key. The

balance displays

p5

.

Step 4 Press

the

PRINT

key. Then the balance displays the function

switches. Set the function table switch and internal mass
correction switch to "

1

" as shown above using the following keys.

SAMPLE

key...To select the switch to change the value.

The selected digit blinks.

RE-ZERO

key ..To change the parameter of the switch selected.





Step 5 Press

the

PRINT

key to store the new setting.

The balance returns to the weighing mode.

Step 6 Press and hold the

SAMPLE

key to enter the function

table and release the key when

ba5fnc

is displayed.

Step 7 Press

the

SAMPLE

key several times until

C5 in

is

displayed, then release the key.

Step 8 Press

the

PRINT

key to enter the procedure for

correcting the internal mass value.

Step 9 Correct the internal mass value using the following keys.

RE-ZERO

key..To select the value. (-50 digits appear

after +50 digits)

PRINT

key.......To store the new value and display the

next menu item of the function table.

Correction the internal

mass by +0.3g at 10kg.

Calibrate with the

corrected internal mass.

The same mass

10kg

10kg

10000.0 g

10000.3 g

Corrected external weight

With these keys Press the key

held down

Press and hold

Function table switch

Correction switch of internal mass