A&D Weighing GF-32K User Manual
Page 32
30
CAL
key..........To cancel the correction and display the next
menu item of the function table.
Step 10 Press the
CAL
key to return the weighing mode.
Step 11 Press the
CAL
key to calibrate the balance using the internal mass.
Step 12 Place the external weight on the pan and confirm that the correction has been
performed properly. In this example, confirm that the value displayed is within the
range that is described at "Accuracy after calibration using the internal mass" of "20.
Specification ". If the value is incorrect, repeat the correction.