beautypg.com

3. percent mode (%), 4. animal weighing mode (hold function) – A&D Weighing GF-32K User Manual

Page 21

background image

19

6-3. Percent Mode (%)

The percent mode displays the weighting value in percentage compared to a 100%

reference mass and is used for target weighing or checking the sample variance.

Selecting The Percent Mode

Step 1 Press the

MODE

key to select the unit

%

(Percent mode). If the percent mode can not be
selected, refer to "5. Weighing Units".

Storing The 100% Reference Mass

Step 2 Press the

SAMPLE

key to enter the 100%

reference mass storing mode.

Even in the storing mode, pressing the

MODE

key will switch to the next mode.

Step 3 Place a container on the weighing pan, if

necessary. Press the

RE-ZERO

key to cancel the

weight (tare). The balance displays

100 0 %

.

Step 4 Place the sample to be set as the 100%

reference mass on the pan or in the container.

Step 5 Press the PRINT key to store the reference

mass. The balance displays

100.00 %

. (The

decimal point position depends on the reference
value. The reference mass stored, even if the
AC adapter is removed, is maintained in
non-volatile memory.)

Note

If the balance judges that the mass of the

sample is too light to be used as a reference, it
displays

lo

.

Step 6 Remove the sample.

Reading The Percentage

Step 7 Place a sample to be compared to the reference

mass on the pan. The displayed percentage is
based on the 100% reference mass.




6-4. Animal Weighing Mode (Hold Function)

This is the mode to weigh a moving object such as an animal, even when the display of the
weighing data fluctuates. The hold function allows the average weight of the animal to be
displayed. To use the hold function, set the function in the function table. Refer to "10.
Function Table" and "10-3. Description Of The Class "Environment, Display" " for details.

Pan

Container

100%

mass

Place sample

Percentage

This manual is related to the following products: