A&D Weighing GF-32K User Manual
Page 38
36
Class
Item and Parameter
Description
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date to be
added, refer to "10-7. Clock and
Calendar Function".
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
0
600
bps
1
1200
bps
2
2400
bps
3
4800
bps
4
9600
bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
type
Data format
5
CSV format
Refer to "10-5. Description of Item
"Data Format".
0
No limit
t-Up
Timeout
1
1 second
Selects the wait time to receive a
command.
0
No output
erCd
AK, Error code
1
Output
AK: ASCII code 06h
0
Not used
5if
Serial
interface
Ct5
CTS, RTS control
1
Used
Controls CTS and RTS.
0
Water temperature
d5 fnc
Density function
ldin
Liquid density input
1
Liquid density
Available only when density mode is
selected
mlt
Programmable-unit (Multi-unit)
Available only when programmable-unit mode is selected.
Refer to "13. Programmable Units" for details".
Unit
Unit
Refer to "5. Weighing Units".
C5 in
Internal mass correction
Displayed only when the internal mass value correction switch is set
to 1. Refer to "8. Calibration".
id
ID number setting
Refer to "11. ID Number And GLP Report".
: Functions for GX-K series. : Factory settings. Digit is a unit of minimum weighing value.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or data added to the weighing data such as time, date and ID number.