beautypg.com

A&D Weighing GF-32K User Manual

Page 43

background image

41

KF format

5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment
can only communicate using this format.

This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data

is not zero or overloaded.

This format outputs the unit only for a stable value.





MT format

5if type 3

A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weighing data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit






NU (numerical) format 5if type 4

This format outputs only numerical data.

This format consists of nine characters excluding the terminator.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.



CSV format

5if type 5

This format separates the data of A&D standard format and the unit by a comma ( , ).
This format outputs the unit even when the data is overloaded.
When the ID number, data number, time and date are added at "Data output (

dout

)" of the

function table, outputs ID number, data number, date, time and weighing data in this order
and separates each item by a comma and treats all the items as one group of data.

Note

GF-K series can not append date and time.





Stable value
Unstable value

Data Unit

Terminator

Stable header
Unstable header
Overload header

Header Data

Unit

Terminator

Data Terminator

ID number Data number

Date

Time

Weighing data

This manual is related to the following products: