Title block and end block, Operation – A&D Weighing GF-32K User Manual
Page 57
55
Title block and end block
When weight values are recorded as GLP data, a "Title block" is inserted at the beginning
and an "End block" is inserted at the end of a group of weight values in the GLP report.
Notes
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
If the data memory function is used, the "Title block" and "End block" can not be output.
Operation
Step 1
With the weighing data displayed, press and hold the
key, until
5tart
is displayed, then release the key. The "Title block" is output.
Step 2
The weighing data is output according to the parameter setting of the data output
mode (
prt
) of the function table.
Step 3
Press and hold the
key until
recend
is displayed, then release the key.
The "End block" is output.
Setting of "info 1"
Setting of "info 2"
AD-8121 printer format
General format
Space, ASCII 20h
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
-----------
A
-
&
-
D
MODEL
-----
GX-30K
S/N
-----
01234567
ID
-------
ABCDEFG
DATE
--
2004/12/31
START
TIME
----
09:30:00
WT
---
+12345.6
--
g
WT
---
+12346.1
--
g
WT
---
+12346.2
--
g
WT
---
+12346.3
--
g
WT
---
+12345.1
--
g
WT
---
+12345.3
--
g
WT
---
+12347.1
--
g
WT
---
+12346.4
--
g
END
TIME
----
10:40:15
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
~~~~~~~~~~~A~&~D
MODEL~~~~~GX-30K
S/N~~~~~01234567
ID~~~~~~~ABCDEFG
DATE
~~~~~~2004/12/31
START
TIME
~~~~~~~~09:30:00
WT~~~+12345.6~~g
WT~~~+12346.1~~g
WT~~~+12346.2~~g
WT~~~+12346.3~~g
WT~~~+12345.1~~g
WT~~~+12345.3~~g
WT~~~+12347.1~~g
WT~~~+12346.4~~g
END
TIME
~~~~~~~~10:40:15
SIGNATURE
----------------
Title block
Manufacturer
Model
Serial number
ID number
Date
Time
Weighing data
Time
Signature
End block