beautypg.com

2. initializing the balance – A&D Weighing GF-32K User Manual

Page 34

background image

32















9-2. Initializing The Balance

This function returns the following parameters to factory settings.

Calibration

data

Function

table

The sample unit mass value (counting mode),

100% reference mass value (percent mode)

The data that is stored in the balance using the data memory function
External calibration weight and target weight value
Function switch settings
Liquid density and temperature in the density mode


Note Be sure to calibrate the balance after initialization.

Step 1 Press the

ON:OFF

key to turn off the display.

Step 2 While pressing and holding the

PRINT

key and

the

SAMPLE

key, press the

ON:OFF

key to

display

p5

.

Step 3 Press the

SAMPLE

key to display

Clr

.

Step 4 Press the

PRINT

key.

To cancel this operation, press the

CAL

key.

Step 5 Press the

RE-ZERO

key to display

Clr go

.

Step 6 Press the

PRINT

key to initialize the balance.

The balance will automatically return to the weighing
mode.

Function table

0

To inhibit changes to the function table.

1

To permit changes to the function table.

Not used

Calibration using an external weight

0

To inhibit calibration using an external weight.

1

To permit calibration using an external weight.

Not used

GF-K series. The display shown left indicates the factory settings.

With these keys Press the key

held down

This manual is related to the following products: