7. Weighing Speed Adjustment / Self Check Function

7-1. Weighing Speed Adjustment

This function detects the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is
installed and sets the following three weighing speeds
(response characteristics) automatically.


The function has three rates as follows:


Changing the weighing speed changes the display refresh rate.

Indicator Parameter Weighing

Speed Stability

Display refresh rate

FAST

Cond 0

MID.

Cond 1

SLOW

Cond 2

Fast response, Sensitive value

Slow response,

Stable value

If the weighing speed is changed as follows:
MID. or SLOW FAST =10 times/second
FAST MID. or SLOW = 5 times/second




Step 1 Press and hold the

MODE

key until

RESPONSE

is

displayed. And then, press the

MODE

key again

quickly.




Step 2 Press the

MODE

key to select a weighing speed.

Either

FAST

,

MID.

or

SLOW

can be selected.




Step 3 After a few seconds of inactivity the balance displays

end

.

Then, it returns to the weighing mode and displays the
updated response indicator. The response indicator
remains displayed for a while.



Note

The weighing speed adjustment can be changed at "Condition (

Cond

)" of

"Environment, Display (ba5fnc)" in the function table. Refer to "10. Function
Table" for details.

To set a refresh rate of 5 times/second when the response rate is

FAST

or

10 times/second when the response rate is

MID.

or

SLOW

, change the

"Display refresh rate (

5pd

)" parameter of "Environment, Display (ba5fnc)" in

the function table.

Weighing speeds

Release and wait

Press again

Press and hold

Release

Select a parameter

with pressing it.