Entering the density directly (

ldin 1

)

The density currently set (unit : g / cm

3

, factory setting : 1.0000g / cm

3

) is displayed.

Use the following keys to change the value.

The range to set the density is 0.0000g / cm

3

to 1.9999g / cm

3

.

RE-ZERO

key...To set the value of the digit selected.

SAMPLE

key ....To select the digit to change the value.

PRINT

key........To store the change, display

end

and return to the density mode.

(Go to "Step 2 Selecting a parameter to set")

CAL

key...........To cancel the change and return to the density mode.

(Go to "Step 2 Selecting a parameter to set")



Outputting the density.

Step 4 After all the necessary parameters have been set, press

the

SAMPLE

key to output the density calculated.

The density is displayed with three decimal places.

Calculated result