35

Class

Item and Parameter

Description

Cp Hi
Upper limit
Cp lo
Lower limit

Refer to
"10-8. Comparator Function"

Displayed when Cp in 0

is selected.

Cp ref
Reference value
Cp lmt
Tolerance

Refer to
"10-8. Comparator Function"

Displayed when Cp in 1

is selected.

0

Key mode

Accepts the PRINT key only when

the display is stable.

1

Auto print mode A
(Reference = zero)

2

Auto print mode B
(Reference = last stable value)

Outputs data when the display is

stable and conditions of ap-p, ap-b

and the reference value are met.

prt
Data output mode

3

Stream mode /
Interval memory mode

With data 0, outputs data

continuously; with data 2, uses

interval memory.

0

Plus only

Displayed value>Reference

1

Minus only

Displayed value

ap-p
Auto print polarity

2

Both

Regardless of displayed value

0

10 digits

1

100 digits

ap-b
Auto print difference

2

1000 digits

Difference between reference value

and displayed value

0

Not used

1

Stores unit mass in counting mode

2

Stores weighing data

3

Stores calibration data

4

Stores comparator settings

data
Data memory

5

Stores tare value

Related items:

prt, int, d-no, 5-td, info

0

Every measurement

1

2 seconds

2

5 seconds

3

10 seconds

4

30 seconds

5

1 minute

6

2 minute

7

5 minute

int
Interval time

8

10 minute

Interval time in the interval memory

mode when using prt 3, data 2

0

No output

d-no
Data number output

1

Output

Refer to "12. DATA MEMORY"

0

No output

1

Time only

2

Date only

5-td
Time/Date output

3

Time and date

Selects whether or not the time

or date is added to the weighing

data. Refer to "10-7. Clock and

Calendar Function" for details.

0

No output

dout
Data output

5-id
ID number output

1

Output

Selects whether or not the ID

number is output.

: Functions for GX-K series. : Factory settings. Digit is a unit of minimum weighing value.

This manual is related to the following products: