Counting Mode Using The ACAI Function

The ACAI is a function that improves the accuracy of the unit mass automatically by

increasing the number of samples as the counting process.
ACAI: Automatic Counting Accuracy Improvement



Step 8 If a few more samples are added, the

processing indicator turns on. To
prevent an error, add three or more.
The processing indicator does not
turn on if overloaded. Try to add the
same number of samples as
displayed.

Step 9 The balance re-calculates the unit

mass while the processing indicator
is blinking. Do not touch the balance
or samples on the pan until the
processing indicator turns off.

Step 10 Counting accuracy is improved when

the processing indicator turns off.

Each time the above operation is

performed, a more accurate unit
mass will be obtained. There is no
definite upper limit of ACAI range for
the number of samples exceeding
100. Try to add the same number of
samples as displayed.

Step 11 Remove all the samples used in

ACAI and proceed with the counting
operation using the improved unit
mass.


Note ACAI will not function on the unit mass entered using the keys, or digital

input mode.


Add a few more

samples.

The mark turns on and off

during calculation.

The mark turns off after the

unit mass is improved.

From step 7

Renew

The mark turns on at

proper range.