beautypg.com

Using the accumulation function, Outputting the value – A&D Weighing GF-32K User Manual

Page 23

background image

21

Using The Accumulation Function

Use the following keys to operate the accumulation function.

MODE

key .......... Displays the weighing data and the total value alternately each time it is

pressed. While the accumulation function is in use, the unit can not be
changed.

RE-ZERO

key ..... Sets the display to zero while the weighing data is displayed.

Deletes the total value while the total value is displayed.

PRINT

key .......... Outputs and adds the weighing data while the weighing data is displayed.

Outputs the total value while the total value is displayed.


Step 1 Press the

RE-ZERO

key to zero the

display.


Step 2

Place a sample on the pan. The weight
value is displayed.


Step 3 Press the

PRINT

key. The weight

value is added to the total and is
output.

The

accumulation

number at the upper

left of the display increases by one.


Step 4 Repeat steps 1 to 3, when

accumulating more data.


Step 5 Press the

MODE

key to display

the total value.

Outputting the value

Step 6 Press the

PRINT

key to output the

total value.


Step 7 Press the

RE-ZERO

key to delete

the total value.






Notes And Displaying Or Outputting An Overloaded Total

The output format depends on the function table setting.
While the accumulation function is in use, the data memory

function is not available.

To disable the accumulation function, set the

"Accumulation function (

add

)" parameter to "

0

".

When the "Data number output (

d-no

)" parameter is set to

"

1

", the accumulation number will be output before the

weighing data.

Output ST,+012345.6

g

Accumulation

number

Accumulation

number

Output TOTAL(N=012)
ST,+012345.6

g

Accumulation value

Data number is above 99

Accumulation data is

overloaded.

Output TOTAL(N=--)
OL,+9999999E+19

g

This manual is related to the following products: