4-1. Smart Range Function

The GX-32K, GF-32K, GX-8K2 and GF-8K2 are equipped with two ranges of "precision

range" of a higher resolution and "standard range" of normal resolution.

The range is switched automatically depending on the value displayed.

Pressing the

RE-ZERO

key allows weighing in the precision range, regardless of the

tare value. (Smart range function)

The range can be fixed to the standard range, by pressing the

SAMPLE

key.

Note

Once the range is switched to the standard range, it will not switch to the precision

range automatically even when the displayed value becomes within the precision
range value. Press the

RE-ZERO

or

SAMPLE

key to use the precision range again.

Example

GX-32K or GF-32K, precision range = 6.1 kg x 0.1g,
standard range = 31 kg x 1 g.


Step 1 Press the

RE-ZERO

key.

The balance will start weighing, using the
precision range.


Step 2 Place a container on the weighing pan.

When the weighing value exceeds

the precision range, the range will be
switched to the standard range.



Step 3 Press the

RE-ZERO

key.

The balance will be switched to the
precision range.



Step 4 Place a sample on the pan.

When the weighing value is within the
precision range, the balance will perform
a weighing using the precision range.



Precision range/standard range value

Precision range

(after

RE-ZERO

key is pressed)

Standard range

GX-32K, GF-32K

Up to

6100.9

g

6101 to

31008

g

GX-8K2, GF-8K2

Up to

2100.09

g

2100.1 to

8100.8

g

Weighing Pan

Precision range

Tare (Container)

Standard range

Precision range

Sample

Precision range

This manual is related to the following products: