4-1. Smart Range Function
The GX-32K, GF-32K, GX-8K2 and GF-8K2 are equipped with two ranges of "precision
range" of a higher resolution and "standard range" of normal resolution.
The range is switched automatically depending on the value displayed.
Pressing the
RE-ZERO
key allows weighing in the precision range, regardless of the
tare value. (Smart range function)
The range can be fixed to the standard range, by pressing the
SAMPLE
key.
Note
Once the range is switched to the standard range, it will not switch to the precision
range automatically even when the displayed value becomes within the precision
range value. Press the
RE-ZERO
or
SAMPLE
key to use the precision range again.
Example
GX-32K or GF-32K, precision range = 6.1 kg x 0.1g,
standard range = 31 kg x 1 g.
Step 1 Press the
RE-ZERO
key.
The balance will start weighing, using the
precision range.
Step 2 Place a container on the weighing pan.
When the weighing value exceeds
the precision range, the range will be
switched to the standard range.
Step 3 Press the
RE-ZERO
key.
The balance will be switched to the
precision range.
Step 4 Place a sample on the pan.
When the weighing value is within the
precision range, the balance will perform
a weighing using the precision range.
Precision range/standard range value
Precision range
(after
RE-ZERO
key is pressed)
Standard range
GX-32K, GF-32K
Up to
6100.9
g
6101 to
31008
g
GX-8K2, GF-8K2
Up to
2100.09
g
2100.1 to
8100.8
g