Specifications (differs by model type), Specifications (all model types are the same) – KEYENCE GP-M Series User Manual
Page 22

22
Specifications (Differs by model type)
Model
GP-M
001
GP-M
010
GP-M
025
GP-M
100
GP-M
250
GP-M
400
Rated
pressure
-100 to
+100
kPa
-0.1 to
+1
MPa
-0.1 to
+2.5
MPa
0 to +10
MPa
0 to +25
MPa
0 to +40
MPa
Possible
display
range
-120.0 to
+120.0
kPa
-0.210 to
+1.110
MPa
-0.360 to
+2.760
MPa
-1.00 to
+11.00
MPa
-2.50 to
+27.50
MPa
-4.00 to
+44.00
MPa
Allowable
pressure 400 kPa 4 MPa 10 MPa 30 MPa 50 MPa 50 MPa
Burst
pressure
1.5 MPa 15 MPa 35 MPa
100
MPa
100
MPa
100
MPa
Display
resolution 0.1 kPa
0.001 MPa
0.01 MPa
Fluid type
Gas or liquid that will not
corrode the liquid contact
part
Liquid that will not corrode
the liquid contact part
Specifications (All model types are the same)
Type of pressure
Gage pressure
Precision
*1
±1.0% of F.S. or less
Repeatability
*2
±0.3% of F.S. or less
Temperature
characteristics
±0.6% of F.S./10°C
Zero-cut pressure
value
±0.5% of F.S.
Connection
diameter
G3/4 (Changes to the R1/8 male, R1/4 male,
R3/8 male, G1/4 female, NPT1/8 male, and
NPT1/4 male option adapters are available.)
Box rotation angle
Maximum 330°