Responsiveness

The time until 100% response of the internal judgment value can be

set from within the range of 3 to 5000 (ms).

The response time of the analog output (90% response) will be an

additional 2 ms from this point.

„

Power Save Mode

When QP is selected, if the key operation does not occur for 15

seconds (Power Save) will be displayed.

The “”at the time of power save display will move from the left to

the right, and the operation indicator will operate normally. To return

the display status to normal, press any key. When there is an error,

regardless of whether power save mode is QP/Q((, the error will be

displayed.

„

Display Renewal Cycle

Set the interval to renew the display contents. When a value is increased,

only for that time the value will continue to be displayed the same. From

after the renewal until the next renewal, the measurement value will be

averaged so as to control fluctuations in the display value. However,

there will be no influence on the control output and the analog output.

„

Display inversion

When PQ is selected: regular display

When ;'5 is selected: The display will be inverted vertically 180°.

Time chart

Reference

Output
judgment

FT/FT

(OFF delay)

OFF

ON

F5/F5

(ON delay)

OFF

ON

Control output
after delay