Error Message List
Appendix-10
LJ-H1W-M-NOA-E
Head setting
The value for delay count exceeds the
setting range.
Set the value between ( ) and ( ).
* The values displayed within the
parentheses ( ) vary depending on the
sampling count.
Set the delay count between 0 and 99,
depending on the sampling count.
Page 5-3
The maximum value for Manual exceeds the
setting range.
Set the value between 1 and 99.
Set the maximum value for Manual of the
laser intensity between 1 and 99.
Page 5-3
The minimum value for Manual exceeds
the setting range.
Set the value between ( ) and ( ).
* The values displayed within the
parentheses ( ) vary depending on the
maximum value for Manual.
Set the minimum value for Manual of the
laser intensity between 1 and 99 and below
the maximum value for Manual of the laser
intensity.
Page 5-3
The number of times for profile alarm
exceeds the setting range.
Set the value between 0 and 255.
Set the number of times for profile alarm
between 0 and 255.
Page 5-3
The value for correction angle exceeds the
setting range.
Set the value between -10
°
and 10
°
.
Set the correction angle between -10
°
and
10
°
.
Page 5-3
The value for height correction exceeds
the setting range.
Set the value between 0 and 10.
Set the height correction between 0x and
10x.
Page 5-3
Profile setup
The amount of movement in the X direction
exceeds the range.
Set the amount of movement in the X direction
within the range for X measurement width.
Page 5-5
The amount of movement in the Z direction
exceeds the range.
Set the amount of movement in the Z
direction within the range for Z measurement
width.
Page 5-5
The value for delay count between the
heads exceeds the setting range.
Set the value between 0 and 1023.
Set the value for delay count between the
heads between 0 and 1023.
Page 5-5
Failed to acquire profile for Head A.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Pages 4-7,
5-5
Failed to acquire profile for Head B.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Pages 4-7,
5-5
Master
registration
Failed to acquire profile for Head A.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Pages 4-7,
5-5
Failed to acquire profile for Head B.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Pages 4-7,
5-5
Item
Error message
Measure
Reference