Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 61
Brookfi eld Engineering Labs., Inc.
Page 61
Manual No. M08-219-B1211
USA and Canada
UL-permission and CSA Certifi cate are required for the mains connecting cable.
The following minimal requirements are valid for the cable:
No. 18 AWG
Type SV or similar
3-phase
The cable must have a nominal rating of at least 10 A.
The main plug of type NEMA 5-15P (15A, 125V) or NEMA 6-15P (15A, 250V) must have
a grounding contact.
Japan
The following cable types and connection values are required in Japan:
All parts of the cables (cable, socket and plug) must have brand name and registration
number according to the Dentori law.
The following minimal requirements apply to the cable:
0.75 mm
2
, 2-phase
Type VCT or VCTF
3-phase
The cable must have a minimum nominal rating of 7 A.
The mains plug must be a 2-pin plug with grounding contact according to the Japanese
Industrial Standard C8303 (15A, 125V).
Other countries
The connectors of the mains connecting cable must be approved and certifi ed by the
responsible authorities in the respective countries. These authorities are:
Australia — EANSW
Great Britain — BSI
Austria — OVE
Italy — IMQ
Belgium — CEBEC
Netherlands — KEMA
Denmark — DEMKO
Norway — NEMKO
Finland — SETI
Sweden — SEMKO
France — UTE
Switzerland — SEV
Germany — VDE
The cable must be three-phase HAR-cable, type HO5VV-F3, with a minimal phase
cross-section of 1.0 mm
2
. The main connecting cable must be permitted for a rating of
at least 10 A and, depending on country, a nominal voltage of 125 V or 250 VAC.
•
•
-
-
-
•
•
•
•
-
-
-
•
•
•
-