Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 42
Brookfi eld Engineering Labs., Inc.
Page 42
Manual No. M08-219-B1211
The input procedure for start and end value number of measurement points and step duration is
repeated for the next step. The procedure is repeated until all steps have been entered.
The program will then prompt the user for a program name.
Edit PROGRAM name:
TEST
Enter the name under which
the program is to be stored
For this example, we will use “TEST” (see Section VI.4: Input of Numerical Values). The
instrument will prompt the user to store the program name:
(2) TEST
Press
Press the
OK
key to store the program.
If any other key besides
OK
is pressed, the entries are abandoned and
those parameters that existed before editing will be preserved.
VI.6.3 Utilities
→ Print Programs
This function will print the parameters of the defi ned programs in memory.
If the printer is not ready for operation when the “Print Programs” function is initiated, the
following error message will be displayed:
ERROR #1
Printer not ready!
If this error message appears, ensure that the printer is ready and contains paper.
Press the
OK
key to try printing again or press the
ST
key to return to the menu.
VI.6.4 Utilities
→ Measuring Systems
This function provides a method to create new measuring systems or change existing measuring
systems.