Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 40
Brookfi eld Engineering Labs., Inc.
Page 40
Manual No. M08-219-B1211
Select program
1) Prog xyz
2) Test
3) NEW
Available programs
contain a name
All free (non-defi ned)
programs are marked NEW
All free programs are initially marked as NEW. To avoid overwriting existing programs, select
NEW as a program to be edited and select a measuring system. After the selection of the program
with the OK key, you will be requested to select a measuring system.
Attention: Any defi ned program MUST ALWAYS be executed with the same measuring system,
otherwise improper results will be calculated.
Select Meas. system:
1) CC40
>
2) CC25
3) CC14
>
After selecting the measurement system, enter the number of steps (number of ramp and straight
line functions). The number of steps can range from 1 to a maximum of 10 steps. In this
example, we need three steps, so the number “01” is changed to “03” - (see Section VI.4: Input
of Numerical Values). The message “Range error” will appear if <1 or >10 is entered.
Nr of steps: 03
Input number of
steps 1 - 10. This
example is 03.
After entering the number of steps, input the type of measurement.
Select Input:
-D[1/s]
-n[rpm]
-Tau[Pa] >
Available Measurement Types
User Defi ned Range
Shear Rate D{s
-1
]
Depends on the measuring system
Speed n[rpm]
0.1 - 1,000 rpm
Shear Stress Tau [Pa]
Depends on the measuring system
Torque M[
‰
]
0 - 999 promille (= 0 to 50 mNm)
Select the type of measurement and press the
OK
key (in our example “D[1/s]”). Now enter the
user defi ned values for each step one after the other.