Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 57
Brookfi eld Engineering Labs., Inc.
Page 57
Manual No. M08-219-B1211
The data in the following table is for plate/plate systems and is valid for a measuring gap of 1
mm. For other plate distances, please calculate K
⋅γ
with the formula:
K
γ
= (2
π/60) * (R/H)
R = radius, H = gap
Table A-4: Plate/Plate Measuring Systems
(consisting of measuring cone and measuring device Cone/plate)
Measuring System
RP3-25
RP3-50
Shear rate range [s
-1
]
1 - 1,309
2 - 2,618
Shear stress range [Pa]
489 - 16,297
61 - 2,040
Viscosity range [Pa•s]
0.373 - 124,500
0.023 - 7,794
Filling volume [mL]
0.5
2.0
Shear rate factor K
⋅γ
[min/s]
1,309
2,6175
Shear stress factor
τ
‰
[Pa]
16,2297
2,037
Radius of measuring plate R
[mm]
12.5
25.0
The given ranges are standard values (not maximum values)
A.2 Error
Messages
The R/S+ Rheometer is user-friendly regarding errors. When errors occur, they are trapped, and the
user is informed on the LCD. The most frequent error messages are explained in this section.
Range Error
Cause:
User’s error at input. The user has tried to input a value that is less than
the allowed minimum value or is more than the allowed maximum value.
Example of
Faulty Input
of a Preset
Value:
Range error
End value:
Min: 0.90
Max: 1032.80
Allowed
minimum value
Name of value out
of range
Allowed maximum
value
What to Do: Enter the value again. Be sure that the new value is within the allowed
value range.