Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 58
Brookfi eld Engineering Labs., Inc.
Page 58
Manual No. M08-219-B1211
Printer Error
Cause:
The printer has been selected as the output device but is not ready for
printing. Possible reasons include:
a] printer cable is not connected to the rheometer
b] paper is out, printer is not online, other printer errors
Example of
Print Error
Message
Error #1
Printer not ready!
Try again
Abort activity;
do not try again
What to Do:
make sure the printer cable is connected to the rheometer
make sure the paper tray has paper
check that the “ready” LED lights of the printer are on
remove the error that was caused at the printer and press the OK
key on the rheometer
press the ST key to abort the process if you cannot fi nd the
error cause
a]
b]
c]
d]
e]
Zero Calibration Error
Cause:
An impermissibly high value was measured during the zero point calibration
of the rheometer.
Example
Message
Error #3
Zero cal. error!
Please retry cal.
What to Do:
press the ST key
ensure the measuring system has been removed from the
rheometer
retry zero point calibration - if the error occurs again, this is an
indication of a fault in the measuring instrument
a]
b]
c]