Brookfi eld Engineering Labs., Inc.

Page 41

Manual No. M08-219-B1211

The following inputs have to be made for each step:

Input Step Nr1
Start[1/s]:0010.00
End[1/s]: 0100.00
Nr of MP: 010

End Value

Start Value

Number of
Measurement Points
(data points)

Minimum and maximum start and end values depend on the selected measuring system for shear
rate (D[s

-1

]) and shear stress (Tau [Pa]). At input, the rheometer checks the start and end values

and indicates the message “Range error” if out of range:

Range error
Start value:
Min: 0.90
Max: 1032.80

Allowed minimum

value

Name of value out of
range

Allowed maximum
value

For example, a range error message would be displayed if a start value for D[s

-1

] is not within the

range of 0.0 - 1032.8 s

-1

when using measuring system CC25.

If the start value, end value and number of measuring points are acceptable, the user will be
prompted for Step Duration:

Input Step Nr. 1
Time[s]: 0060

Number of steps

being edited

Step duration

Minimum step duration:

Available Measurement Types User Defi ned Range

Shear Rate D{s

-1

]

tmin = number of measuring points * 4 s

Speed n[rpm]

tmin = number of measuring points * 4 s

Shear Stress Tau [Pa]

tmin = number of measuring points * 1 s

Torque M[

]

tmin = number of measuring points * 1 s

Maximum step duration: 3600 s

The instrument will automatically check the input. If there is a range error, “Range error” would
be displayed together with the allowable range.

The more time between two measuring points, the higher the accuracy
of the determined physical parameters!