Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 36
Brookfi eld Engineering Labs., Inc.
Page 36
Manual No. M08-219-B1211
This message will be displayed until the fi rst measuring point is displayed:
100.5
‰
000.90 1/s
Eta: 1.728 Pas
Tau: 1.572 Pa
05s 20.7°C
#01
Torque
Duration
Temperature
Actual pre-set value
Measurement result
Current step
If the torque is less than 30‰, the results may be out of instrument range. The user should consider
changing the user defi ned values to get a torque higher than 100‰.
If the display fi eld for the torque indicates: “Mlow!”, these values are below the range of resolution
of the rheometer.
If the temperature is not displayed, the measuring sensor is not connected and the printout of the
temperature will be printed as “---”.
The display is updated with every new measurement point. The current measurement can be
cancelled at any time with the
ST
key.
After a measurement, or after a stop, the display fi eld for step indicates “END” or “BREAK”.
The LCD alternates at intervals of about four seconds between the last displayed measurement
point and information on the measurement:
Program end
Single measurement
Total time: 200s
Total #MP: 40
Reason for
break or end
Duration until
break or end
Type/name of
executed program
Number of
measurement points
Pressing the
OK
key ends the alternating display and returns to the main menu.
VI.5.4 MAIN
Menu
→ Remote
The “Remote” function initiates measurements to be made under PC control. In this mode, all
functions of the R/S+ Rheometer are controlled by a PC. For PC-controlled measurements, you
need the Rheo 3000 software. This software operates under Microsoft Windows 7™, Windows
2000™, Windows XP™, and Vista™. More detailed information on Rheo 3000 software can be
obtained from Brookfi eld or an authorized dealer.
After selecting the Remote option, the rheometer displays the following:
Remote
Wait for RS232...